Sensitivity Setting in IRD8800 module M88200
Module M88200 of IRD8800 series allows Sensitivity setting in mV/g and mV/mm/sec.
When Sensitivity Unit is set to mV/g, then the card allows setting a sensitivity value between 000.0 and 999.9 mV/g. Typically, the factory set value is kept at 100.0 mV/g to match the standard accelerometers available in the Market.
When Sensitivity Unit is set to mV/mm/sec, then the card allows setting a sensitivity value between 0.000 and 9.999 mV/mm/sec. Typically, the factory set value is kept at 4.000 mV/mm/sec to match the standard velocity output accelerometers (velometers) available in the Market.
Related Articles
Sensitivity options in IRD8800 Module M88210
The Relative Shaft Vibration Module M88210 of IRD8800 Series allows different Sensitivity options to accommodate different Eddy Current Probes of different manufacturers keeping different Sensitivities. The option to minutely adjust sensitivity also ...
IRD8800 Module M88300 Menu Structure
Attached to this article is the menu structure of Module M88300 of IRD8800 Series
Sensitivity of M88280
M88280 Module of IRD8800 Series is meant for Vibration Monitoring of Generator End Winding. The sensitivity of the module can be adjusted based on the input high temperature sensor and Charge Amplifier. The Sensitivity for a Charge Amplifier with an ...
Calibration Check Table for IRD8800 Series Module M88200
Pre-requisites If you wish to check the Calibration of IRD8800 Series Module M88200, then you would need the following tools: A Signal Generator with the capability to generate a Sine Wave of a fixed frequency and fixed amplitude and an accuracy of ...
Calibration Check Table for IRD8800 Series Module M88210
Pre-requisites If you wish to check the Calibration of IRD8800 Series Module M88210, then you would need the following tools: A Signal Generator with the capability to generate a Sine Wave of a fixed frequency and fixed amplitude and an accuracy of ...