Sensitivity options in IRD8800 Module M88210
The Relative Shaft Vibration Module M88210 of IRD8800 Series allows different Sensitivity options to accommodate different Eddy Current Probes of different manufacturers keeping different Sensitivities. The option to minutely adjust sensitivity also allows for Pair Calibration and elimination of Error in Measurement when connected to Non- Contact Eddy Current Pickup (that have deviation from their nominal sensitivity). The options in M88210 are:
- 8 mV/micron - upon selecting this option, the user can then fine tune sensitivity to a value between 6.400 mV/micron to 9.600 mV/micron
- 4 mV/micron - upon selecting this option, the user can then fine tune sensitivity to a value between 3.200 mV/micron and 4.800 mV/micron
Related Articles
Sensitivity Options in IRD8800 Module M88250
IRD8800 Series is an API 670 Compliant Machinery Protection System (MPS). It's module M88250 is used for Casing Vibration Monitoring and Protection paired with IRD544M Sensor. IRD544M sensors are designed for Nominal Sensitivity 42.52 mV/mm/sec +/- ...
User Manual of IRD6600 (Entek IRD 6600)
IRD6600 is a now Obsolete system that has been replaced by IRD8800. User manual of IRD6600 is attached to this Article.
Sensitivity Setting in IRD8800 module M88200
Module M88200 of IRD8800 series allows Sensitivity setting in mV/g and mV/mm/sec. M88200 can accept Sensors with Sensitivity 10 mV/g, 50 mV/g, 100 mV/g, 200 mV/g and 500 mV/g (all commonly used sensitivities in the market) The Full Scale possible ...
Sensitivity of M88280
M88280 Module of IRD8800 Series is meant for Vibration Monitoring of Generator End Winding. The sensitivity of the module can be adjusted based on the input high temperature sensor and Charge Amplifier. The Sensitivity for a Charge Amplifier with an ...
IRD7100 Speed Monitoring Options
IRD7100 has the following Speed Monitoring Options. Units are Pulses/Second. Pulses are multiplication of Machine RPM and No. of Tooth 8 Pulses/Second 16 Pulses/Second 20 Pulses/Second 63.25 Pulses/Second 200 Pulses/Second 632.5 Pulses/Second 2000 ...