Sensitivity Options in M88210 Module of IRD8800 Series

Sensitivity options in IRD8800 Module M88210

The Relative Shaft Vibration Module M88210 of IRD8800 Series allows different Sensitivity options to accommodate different Eddy Current Probes of different manufacturers keeping different Sensitivities. The option to minutely adjust sensitivity also allows for Pair Calibration and elimination of Error in Measurement when connected to Non- Contact Eddy Current Pickup (that have deviation from their nominal sensitivity). The options in M88210 are:
  1. 8 mV/micron - upon selecting this option, the user can then fine tune sensitivity to a value between 6.400 mV/micron to 9.600 mV/micron
  2. 4 mV/micron - upon selecting this option, the user can then fine tune sensitivity to a value between 3.200 mV/micron and 4.800 mV/micron

    • Related Articles

    • Sensitivity Setting in IRD8800 module M88200

      Module M88200 of IRD8800 series allows Sensitivity setting in mV/g and mV/mm/sec. When Sensitivity Unit is set to mV/g, then the card allows setting a sensitivity value between 000.0 and 999.9 mV/g. Typically, the factory set value is kept at 100.0 ...
    • IRD8800 Module M88300 Menu Structure

      Attached to this article is the menu structure of Module M88300 of IRD8800 Series
    • Sensitivity of M88280

      M88280 Module of IRD8800 Series is meant for Vibration Monitoring of Generator End Winding. The sensitivity of the module can be adjusted based on the input high temperature sensor and Charge Amplifier. The Sensitivity for a Charge Amplifier with an ...
    • IRD7100 Speed Monitoring Options

      IRD7100 has the following Speed Monitoring Options. Units are Pulses/Second. Pulses are multiplication of Machine RPM and No. of Tooth 8 Pulses/Second 16 Pulses/Second 20 Pulses/Second 63.25 Pulses/Second 200 Pulses/Second 632.5 Pulses/Second 2000 ...
    • Calibration Check Table for IRD8800 Series Module M88200

      Pre-requisites If you wish to check the Calibration of IRD8800 Series Module M88200, then you would need the following tools: A Signal Generator with the capability to generate a Sine Wave of a fixed frequency and fixed amplitude and an accuracy of ...