IRD dataPAC2040 User Manual

IRD dataPAC2040 User Manual

IRD dataPAC2040 User Manual

4-Channel Portable Vibration Analyser & Data Collector

November 2026

IRD Mechanalysis Limited · www.irdmech.com · sales@irdmech.com

1. About this manual

This manual is for people who use dataPAC2040 in the field or on the bench: route-based condition monitoring, machinery diagnostics and field balancing.

The product name is written dataPAC2040, with no space, using IRD’s historic “dataPAC” spelling. A two-channel model, dataPAC2020, uses the same instrument hardware. That model is covered in section 11.

2. Safety

  • Use the instrument only for machinery vibration measurement, data collection and field balancing, as described here.
  • Follow site safety rules, including lock-out, permit-to-work, and any rule about working on rotating or electrically live machines.
  • Keep clear of rotating shafts, couplings, belts and other moving parts when placing sensors or applying balancing weights.
  • Do not open the instrument except as described for a field-swappable battery. There are no other user-serviceable parts inside.
  • Fit the tethered caps on unused connectors. Ingress protection (IP65) applies with the caps fitted. With cables connected, sealing is lower.
  • Use only the USB-C charging arrangement supplied or approved by IRD. The battery is a rechargeable lithium-ion pack. Do not crush, short, or expose it to fire or high heat. Charge and store within the stated temperature range.
  • The instrument is not approved for hazardous areas (ATEX / IECEx). Do not take it into a zoned hazardous area.
  • If the housing, connectors, screen or battery are damaged, stop using the instrument and contact IRD.

For repair and calibration, IRD provides National Traceable calibration. There is a three-month warranty on repairs.

3. Product overview

dataPAC2040 is a compact, rugged four-channel analyser and data collector. It acquires four dynamic channels at the same time, plus a dedicated tachometer / phase-reference input.

It is used for:

  • Route data collection — overall vibration, spectrum, time waveform and gSE (spike energy) captured together in one record, with an on-screen alarm colour.
  • Diagnostics — spectra, waveforms, phase, orbits, cross-channel measurements, bump tests, run-up and coast-down, order tracking, and related tools.
  • Field balancing — single- and two-plane balancing, using up to four sensors at once.

The instrument has a 7-inch colour IPS touchscreen in landscape orientation, sealed function keys below the display, and a protective elastomer impact boot. The display is sunlight-readable, optically bonded and glove-capable. All sensor connectors sit on the top edge and point upward, so cables hang clear of the screen, keys and hands. A USB-C port on the lid is used for charging and data.

A companion desktop application for Microsoft Windows builds routes, reviews trends and alarms, analyses recorded data, and produces reports.

4. Controls and connectors

4.1 Display and keys

  • 7-inch colour IPS touchscreen, landscape.
  • Sealed keys below the display: F1, F2, F3, F4 and HOME.

Route measurements can be taken with the keys, without needing the touchscreen.

4.2 Connectors

All vibration and tachometer connectors are sealed panel-mount parts on the top edge of the housing, at a 25 mm pitch, in this order from left to right:

PositionMarkingConnectorUse
1CH1Sealed BNCVibration (or other dynamic) input. This channel is BNC only.
2CH2Sealed BNCVibration input. Also used as the X axis when a triaxial sensor is connected.
3CH3Sealed BNCVibration input. Also used as the Y axis when a triaxial sensor is connected.
4CH4Sealed BNCVibration input. Also used as the Z axis when a triaxial sensor is connected.
5TRIAXM12, 8-pin, A-coded, femaleTriaxial accelerometer. Uses CH2, CH3 and CH4.
6TACHOM12, 4-pin, A-coded, femaleTachometer / phase reference. Isolated input.

USB-C is on the lid, with a tethered sealed cap. It is used for charging and for transferring data.

Colour coding: CH1 blue, CH2 yellow, CH3 green, CH4 red, TACHO white, on the connector, the cable boot and the trace on the screen.

4.3 Using the BNC and triaxial ports together

  • CH1 is always the CH1 BNC. It is not part of the triaxial input.
  • When a triaxial sensor is connected to the M12 triaxial port, it occupies CH2, CH3 and CH4. Do not also connect BNC sensors to CH2, CH3 or CH4 at the same time.
  • When the triaxial port is not in use, CH2, CH3 and CH4 are used from their BNCs.

5. Connecting sensors and the tachometer

5.1 Vibration sensors

The instrument can power and read:

  • IEPE / constant-current accelerometers, with selectable 0, 2 or 4 mA at 24 V, and a bias-voltage and cable-fault check.
  • AC voltage signals, ±10 V.
  • DC voltage, including proximity-probe gap, 0 to −24 V and ±24 V.
  • 4–20 mA process signals.
  • Velocity-coil transducers.

Place the sensor on a clean, firm measurement point, with a magnet or a stud as the measurement method requires. Wait for IEPE bias to settle before recording. If the instrument reports a cable or bias fault, check the cable, the connector and the sensor before recording.

5.2 Tachometer / phase reference

The TACHO port is a dedicated, optically isolated input.

  • Auto or adjustable trigger.
  • Speed range 1 to 300,000 RPM.
  • A 5 V supply is available for a laser tachometer.
  • A strobe-trigger output is provided.

Use this input for phase, order-based measurements, orbits that need a reference, run-up and coast-down, and field balancing. Confirm the trigger on a stable pulse before relying on phase.

5.3 Caps and cables

Fit caps on unused ports. Use IRD or compatible cables with adequate strain relief. A safety-disconnect or swivel cable, if used, plugs into the same panel connectors.

6. Taking measurements

The analyser uses 24-bit analogue-to-digital conversion, with simultaneous sampling on all four channels, and a dynamic range of at least 110 dB.

6.1 Frequency range and resolution

  • Frequency range (Fmax): up to 80 kHz on one channel; up to 40 kHz with all four channels active.
  • FFT resolution: up to 25,600 lines on one channel; up to 12,800 lines on all four channels at once. Zoom FFT is available.
  • Windows: Hanning, flat-top and rectangular.
  • Averaging: linear, exponential, peak-hold and synchronous.

6.2 Measured quantities

Acceleration, velocity and displacement. Overall vibration can be assessed against ISO 20816.

6.3 Core functions

  • Overall level.
  • Frequency spectrum.
  • Time waveform.
  • gSE (spike energy) bearing demodulation — IRD’s bearing-demodulation measurement, including a gSE spectrum with bearing fault-frequency markers.
  • Multi-band envelope demodulation.

6.4 Phase, orbit and cross-channel

  • Phase at 1× to n× running speed.
  • Orbit.
  • Cross-channel phase, coherence and frequency-response (FRF).

6.5 Advanced diagnostics

  • Bump / impact test.
  • Run-up and coast-down, with Bode, Nyquist and waterfall plots.
  • Order tracking.
  • Time-synchronous averaging (TSA).

6.6 Octave and 1/3-octave bands

Octave and 1/3-octave band analysis, 1 Hz to 20 kHz, linear or A-weighted, designed to IEC 61260-1 Class 1.

6.7 User-defined frequency bands

Up to 12 bands can be set on each measurement point, in hertz or in orders of running speed. Each band has its own overall level and its own alert and danger limits, and is trended in the PC software.

7. Route data collection

Routes are built on the companion PC software (plant tree: site, area, machine, point), with sensors, measurement templates and alarm limits, then loaded to the instrument by USB-C or Wi-Fi.

In the field:

  1. The instrument presents the next measurement point.
  2. Place the sensor. The instrument checks bias and settling.
  3. Start a recording. Overall, spectrum, waveform and gSE are captured together in one record. The result is shown as a level with an alarm colour.
  4. The route moves to the next point. A recording can be repeated or taken again if needed.

Notes and a reason for skipping a point can be recorded at the point. After the round, upload the data by USB-C or Wi-Fi. The PC software merges the measurements into the plant database, updates trends, and raises exceptions.

A standard route point is recorded in a single action, using the keys, and can be done with gloves on.

8. Field balancing

dataPAC2040 supports single-plane and two-plane field balancing, using up to four sensors at once, over 30 to 60,000 RPM.

Functions:

  • Trial-weight calculation.
  • Stored influence coefficients, so a repeat job on a known machine can be a one-shot run.
  • Trim runs.
  • Split or combine correction weights to fixed positions or blades.
  • A tolerance check to ISO 21940-11.
  • A balancing report from the PC software.

Always observe site rules when stopping a machine, fitting trial weights, and running it up. Follow the on-instrument balancing steps.

9. Companion PC software

A desktop application for Microsoft Windows accompanies the instrument.

Functions:

  • Plant tree and route builder.
  • Download of routes and upload of measurements, by USB-C or Wi-Fi.
  • Trend and alarm-exception review.
  • Spectrum and waveform analysis, including cursors, harmonics, sidebands, waterfall, orbit, and Bode / Nyquist plots.
  • Bearing fault-frequency overlay from a bearing database.
  • Balancing jobs and reports.
  • Report output in PDF, Word and Excel.
  • Export to CSV and UFF58. Recorded data are kept in a published package format so they are not locked to a single application.

Wi-Fi operates at 2.4 GHz and 5 GHz. USB-C supports both device and host operation.

10. Charging, storage and care

  • Battery: rechargeable lithium-ion pack, which can be swapped in the field.
  • Charging: USB-C, including USB Power Delivery.
  • Operating time: at least 10 hours of typical route use on one charge.
  • Local storage: internal memory of 8 GB or greater, plus microSD expansion.
  • Operating temperature: −10 to +50 °C.
  • Ingress protection: IP65 with connector caps fitted.
  • Drop resistance: 1.5 m onto concrete.
  • Keep the screen, keys and connectors clean and dry. Do not use solvents that can damage the boot or the seals.
  • Store the instrument in its carrying case, with caps fitted, and with a partial charge if it will be unused for a long period.
  • The instrument is supplied in the protective impact boot. Leave the boot on in the field.

11. dataPAC2020 and the upgrade to dataPAC2040

dataPAC2020 is the two-channel model. It uses the same hardware as dataPAC2040: the same enclosure, the same front panel, and the same electronics.

On a dataPAC2020, the instrument enables CH1, CH2 and the tachometer, and locks CH3, CH4 and the triaxial port. The locked ports remain on the panel, with a dataPAC2020 nameplate, so that a later upgrade does not need a workshop visit or a board change. The 2020 kit can be supplied without a triaxial sensor.

A customer who later wants four channels and the triaxial port can purchase the dataPAC2020 to dataPAC2040 Upgrade. IRD then unlocks the instrument to a dataPAC2040. The licence is tied to the instrument’s serial number. There is no return to the factory for the unlock itself.

If a port is locked, the instrument shows that the port can be opened by that upgrade, rather than appearing as a fault.

12. Technical specifications

Specifications are subject to change without notice.

Inputs

Dynamic channels4, acquired simultaneously
Tachometer / phase reference1 dedicated isolated input; auto or adjustable trigger; 1 to 300,000 RPM; 5 V supply for a laser tachometer; strobe-trigger output
Sensor powerIEPE / constant-current, selectable 0 / 2 / 4 mA at 24 V, with bias-voltage and cable-fault check
Other signal typesAC voltage (±10 V); DC voltage including proximity-probe gap (0 to −24 V and ±24 V); 4–20 mA; velocity-coil transducers
ConnectorsFour sealed BNCs (CH1–CH4); one M12 8-pin triaxial port; one M12 4-pin tachometer port; USB-C on the lid with a tethered sealed cap. Sensor and tacho connectors are on the top edge and point upward.

Measurement and analysis

Analogue-to-digital conversion24-bit, simultaneous sampling on all four channels. Dynamic range ≥ 110 dB
Frequency range (Fmax)Up to 80 kHz on one channel; up to 40 kHz with all four channels active
FFT resolutionUp to 25,600 lines on one channel; up to 12,800 lines on all four channels simultaneously. Zoom FFT
Measured quantitiesAcceleration, velocity and displacement. Overall vibration to ISO 20816
Core functionsOverall level, frequency spectrum, time waveform, gSE (spike energy) bearing demodulation, multi-band envelope demodulation
Phase, orbit and cross-channelPhase at 1× to n× running speed; orbit; cross-channel phase, coherence and FRF
Advanced diagnosticsBump / impact test; run-up and coast-down with Bode, Nyquist and waterfall plots; order tracking; time-synchronous averaging (TSA)
Route data collectionOverall, spectrum, waveform and gSE captured together in one record, with on-screen alarm colour
Field balancingSingle- and two-plane, using up to four sensors at once; 30 to 60,000 RPM; trial-weight calculation; stored influence coefficients; trim; split or combine correction weights; ISO 21940-11 tolerance check
Windows and averagingHanning, flat-top and rectangular windows; linear, exponential, peak-hold and synchronous averaging
Octave and 1/3-octave1 Hz to 20 kHz, linear or A-weighted (designed to IEC 61260-1 Class 1)
User-defined frequency bandsUp to 12 per measurement point, in Hz or in orders of running speed, each with alert and danger limits, trended in the PC software

Display, data and power

Display7-inch colour IPS touchscreen, landscape, sunlight-readable, optically bonded, glove-capable
KeysSealed F1, F2, F3, F4 and HOME, below the display
Local storageInternal memory (8 GB or greater) plus microSD expansion
Wired interfaceUSB-C for charging and data. Device and host operation
WirelessWi-Fi, 2.4 GHz and 5 GHz
BatteryRechargeable lithium-ion pack, field-swappable
ChargingUSB-C, including USB Power Delivery
Operating time≥ 10 hours typical route use on one charge

Mechanical and environmental

EnclosureHandheld industrial housing with a protective elastomer impact boot. Dark housing with a blue boot
Dimensions215 × 150 × 46 mm (housing). 221 × 156 × 54.3 mm with the impact boot fitted
Weight≤ 1.1 kg including battery
Ingress protectionIP65 with connector caps fitted
Operating temperature−10 to +50 °C
Drop resistance1.5 m onto concrete
ComplianceDesigned for EMC (EN 61326-1), electrical safety (IEC 61010-1) and rechargeable-battery safety and transport (IEC 62133 and UN 38.3, with applicable national registration).

13. Support

For sales, application advice, repair and calibration:

IRD Mechanalysis Limited

www.irdmech.com

sales@irdmech.com

Quote the instrument’s serial number in every support request.