4-Channel Portable Vibration Analyser & Data Collector
November 2026
IRD Mechanalysis Limited · www.irdmech.com · sales@irdmech.com
This manual is for people who use dataPAC2040 in the field or on the bench: route-based condition monitoring, machinery diagnostics and field balancing.
The product name is written dataPAC2040, with no space, using IRD’s historic “dataPAC” spelling. A two-channel model, dataPAC2020, uses the same instrument hardware. That model is covered in section 11.
For repair and calibration, IRD provides National Traceable calibration. There is a three-month warranty on repairs.
dataPAC2040 is a compact, rugged four-channel analyser and data collector. It acquires four dynamic channels at the same time, plus a dedicated tachometer / phase-reference input.
It is used for:
The instrument has a 7-inch colour IPS touchscreen in landscape orientation, sealed function keys below the display, and a protective elastomer impact boot. The display is sunlight-readable, optically bonded and glove-capable. All sensor connectors sit on the top edge and point upward, so cables hang clear of the screen, keys and hands. A USB-C port on the lid is used for charging and data.
A companion desktop application for Microsoft Windows builds routes, reviews trends and alarms, analyses recorded data, and produces reports.
Route measurements can be taken with the keys, without needing the touchscreen.
All vibration and tachometer connectors are sealed panel-mount parts on the top edge of the housing, at a 25 mm pitch, in this order from left to right:
| Position | Marking | Connector | Use |
|---|---|---|---|
| 1 | CH1 | Sealed BNC | Vibration (or other dynamic) input. This channel is BNC only. |
| 2 | CH2 | Sealed BNC | Vibration input. Also used as the X axis when a triaxial sensor is connected. |
| 3 | CH3 | Sealed BNC | Vibration input. Also used as the Y axis when a triaxial sensor is connected. |
| 4 | CH4 | Sealed BNC | Vibration input. Also used as the Z axis when a triaxial sensor is connected. |
| 5 | TRIAX | M12, 8-pin, A-coded, female | Triaxial accelerometer. Uses CH2, CH3 and CH4. |
| 6 | TACHO | M12, 4-pin, A-coded, female | Tachometer / phase reference. Isolated input. |
USB-C is on the lid, with a tethered sealed cap. It is used for charging and for transferring data.
Colour coding: CH1 blue, CH2 yellow, CH3 green, CH4 red, TACHO white, on the connector, the cable boot and the trace on the screen.
The instrument can power and read:
Place the sensor on a clean, firm measurement point, with a magnet or a stud as the measurement method requires. Wait for IEPE bias to settle before recording. If the instrument reports a cable or bias fault, check the cable, the connector and the sensor before recording.
The TACHO port is a dedicated, optically isolated input.
Use this input for phase, order-based measurements, orbits that need a reference, run-up and coast-down, and field balancing. Confirm the trigger on a stable pulse before relying on phase.
Fit caps on unused ports. Use IRD or compatible cables with adequate strain relief. A safety-disconnect or swivel cable, if used, plugs into the same panel connectors.
The analyser uses 24-bit analogue-to-digital conversion, with simultaneous sampling on all four channels, and a dynamic range of at least 110 dB.
Acceleration, velocity and displacement. Overall vibration can be assessed against ISO 20816.
Octave and 1/3-octave band analysis, 1 Hz to 20 kHz, linear or A-weighted, designed to IEC 61260-1 Class 1.
Up to 12 bands can be set on each measurement point, in hertz or in orders of running speed. Each band has its own overall level and its own alert and danger limits, and is trended in the PC software.
Routes are built on the companion PC software (plant tree: site, area, machine, point), with sensors, measurement templates and alarm limits, then loaded to the instrument by USB-C or Wi-Fi.
In the field:
Notes and a reason for skipping a point can be recorded at the point. After the round, upload the data by USB-C or Wi-Fi. The PC software merges the measurements into the plant database, updates trends, and raises exceptions.
A standard route point is recorded in a single action, using the keys, and can be done with gloves on.
dataPAC2040 supports single-plane and two-plane field balancing, using up to four sensors at once, over 30 to 60,000 RPM.
Functions:
Always observe site rules when stopping a machine, fitting trial weights, and running it up. Follow the on-instrument balancing steps.
A desktop application for Microsoft Windows accompanies the instrument.
Functions:
Wi-Fi operates at 2.4 GHz and 5 GHz. USB-C supports both device and host operation.
dataPAC2020 is the two-channel model. It uses the same hardware as dataPAC2040: the same enclosure, the same front panel, and the same electronics.
On a dataPAC2020, the instrument enables CH1, CH2 and the tachometer, and locks CH3, CH4 and the triaxial port. The locked ports remain on the panel, with a dataPAC2020 nameplate, so that a later upgrade does not need a workshop visit or a board change. The 2020 kit can be supplied without a triaxial sensor.
A customer who later wants four channels and the triaxial port can purchase the dataPAC2020 to dataPAC2040 Upgrade. IRD then unlocks the instrument to a dataPAC2040. The licence is tied to the instrument’s serial number. There is no return to the factory for the unlock itself.
If a port is locked, the instrument shows that the port can be opened by that upgrade, rather than appearing as a fault.
Specifications are subject to change without notice.
| Dynamic channels | 4, acquired simultaneously |
| Tachometer / phase reference | 1 dedicated isolated input; auto or adjustable trigger; 1 to 300,000 RPM; 5 V supply for a laser tachometer; strobe-trigger output |
| Sensor power | IEPE / constant-current, selectable 0 / 2 / 4 mA at 24 V, with bias-voltage and cable-fault check |
| Other signal types | AC voltage (±10 V); DC voltage including proximity-probe gap (0 to −24 V and ±24 V); 4–20 mA; velocity-coil transducers |
| Connectors | Four sealed BNCs (CH1–CH4); one M12 8-pin triaxial port; one M12 4-pin tachometer port; USB-C on the lid with a tethered sealed cap. Sensor and tacho connectors are on the top edge and point upward. |
| Analogue-to-digital conversion | 24-bit, simultaneous sampling on all four channels. Dynamic range ≥ 110 dB |
| Frequency range (Fmax) | Up to 80 kHz on one channel; up to 40 kHz with all four channels active |
| FFT resolution | Up to 25,600 lines on one channel; up to 12,800 lines on all four channels simultaneously. Zoom FFT |
| Measured quantities | Acceleration, velocity and displacement. Overall vibration to ISO 20816 |
| Core functions | Overall level, frequency spectrum, time waveform, gSE (spike energy) bearing demodulation, multi-band envelope demodulation |
| Phase, orbit and cross-channel | Phase at 1× to n× running speed; orbit; cross-channel phase, coherence and FRF |
| Advanced diagnostics | Bump / impact test; run-up and coast-down with Bode, Nyquist and waterfall plots; order tracking; time-synchronous averaging (TSA) |
| Route data collection | Overall, spectrum, waveform and gSE captured together in one record, with on-screen alarm colour |
| Field balancing | Single- and two-plane, using up to four sensors at once; 30 to 60,000 RPM; trial-weight calculation; stored influence coefficients; trim; split or combine correction weights; ISO 21940-11 tolerance check |
| Windows and averaging | Hanning, flat-top and rectangular windows; linear, exponential, peak-hold and synchronous averaging |
| Octave and 1/3-octave | 1 Hz to 20 kHz, linear or A-weighted (designed to IEC 61260-1 Class 1) |
| User-defined frequency bands | Up to 12 per measurement point, in Hz or in orders of running speed, each with alert and danger limits, trended in the PC software |
| Display | 7-inch colour IPS touchscreen, landscape, sunlight-readable, optically bonded, glove-capable |
| Keys | Sealed F1, F2, F3, F4 and HOME, below the display |
| Local storage | Internal memory (8 GB or greater) plus microSD expansion |
| Wired interface | USB-C for charging and data. Device and host operation |
| Wireless | Wi-Fi, 2.4 GHz and 5 GHz |
| Battery | Rechargeable lithium-ion pack, field-swappable |
| Charging | USB-C, including USB Power Delivery |
| Operating time | ≥ 10 hours typical route use on one charge |
| Enclosure | Handheld industrial housing with a protective elastomer impact boot. Dark housing with a blue boot |
| Dimensions | 215 × 150 × 46 mm (housing). 221 × 156 × 54.3 mm with the impact boot fitted |
| Weight | ≤ 1.1 kg including battery |
| Ingress protection | IP65 with connector caps fitted |
| Operating temperature | −10 to +50 °C |
| Drop resistance | 1.5 m onto concrete |
| Compliance | Designed for EMC (EN 61326-1), electrical safety (IEC 61010-1) and rechargeable-battery safety and transport (IEC 62133 and UN 38.3, with applicable national registration). |
For sales, application advice, repair and calibration:
IRD Mechanalysis Limited