If the Calibrator has the facility to change the frequency as well as amplitude, then it is even better and will allow for checking at multiple points.
To check the Calibration, connect the Meter to its Sensor. Place the Sensor on the Calibrator. Give an input vibration of Frequency 100 Hz and Amplitude 1 g peak. The meter should then read the following:
- In Velocity Mode, the meter should read 11 mm/sec RMS +/- 10%
- In Displacement Mode, the meter should read 50 microns peak-peak +/- 10%